Description
Maquet Datascope CS 100i
The Datascope IAPB can be used in an operating room, catheter lab, or critical care setting. The Datascope can be customized to fit each need.
- Automatically evaluates and selects optimal trigger source and lead
- Sets inflation at the dicrotic notch and deflates IAB at the start of systole
- Responds to changes in signal quality by selecting new trigger source
Automated trigger source and timing management provide continuous, consistent support for your patient.
Maquet Datascope Cs 100i Specifications
Dimensions
-
- Height: 30 in (76.2 cm)
- Width: 22.5 in (57.2 cm)
- Depth: 13.2 in (33.5 cm)
- On Cart
- Height: 42.2 in (114.8 cm)
- Width: 17.6 in (44.7 cm)
- Depth: 23.4 in ( 59.4 cm)
- Off Cart
- Height: 28.2 in (71.6 cm)
- Width: 11.3 in (28.7 cm)
- Depth: 21.5 in ( 54.6 cm)
- Monitor weight: 9.5 lb (4.3 kg)
- Console weight: 133.2 lb ( 60.4 kg)
Color Display
- Width:8.3 in (21cm)
- Height: 6.2 in (15.8 cm)
- viewing angle:
- Up: 45°
- Down 55°
- Right 70°
- Left 70°
- Rotates 330°
- Tilts180°
- Detachable Laptop-like closure for storage and protection
Power
- Mains Voltage: 100-120 VAC ± 10% or 220-240 VAC ± 10%
- Mains Frequency: 50/60 Hz ± 3 Hz
- Internal Battery: 24 VDC (nominal), 17.2 Amp-hour, 3 hrs.(approx. @ 90 bpm)
- Battery Type: Maintenance-free; Sealed lead-acid
General information
- ECG Gain (default): 1 V output per 1 mV input ±5% (waveform automatically scaled to occupy ECG display window)
- Gain (variable): 0.15 to 3.0 cm/mV ±20% (autoscaling disabled)
- Frequency Response: 0.5-12 Hz (display); 0.5-135 Hz (Output to Monitor)
- Defibrillator Protection: Discharge levels 360 J (trace returns to the screen in 5 sec. max)
Operating Conditions
- Operating Temperature: 10°C – 40°C
- Operating Humidity: 5 – 95% (R.H.) non-condensing
- Operating Altitude: 0 – 12,000 feet (3,657 m); automatic altitude correction for IAB pressure