Description
Covidien Valleylab FX-C Features
The Covidien Force FX-C is an electrosurgical generator, also referred to as the “Covidien Force FX electrosurgical generator C ” , that provides isolated output power for cutting, desiccating, and fulgurating tissue during bipolar and monopolar surgery.
- Instant Respons technology ensures that the power delivered remains virtually constant, regardless of the tissue type
- Improved performance at lower power settings minimizes the risk of tissue damage and neuromuscular stimulation
- Three internal microcontrollers reduce system reaction time and increase the system’s processing speed
- Spray coagulation voltage of no more than 9000 volts peak-to-peak output for broad, but superficial coagulation with limited capacitive coupling
- A Power Efficiency Rating (PER) of approximately 98 for accurate and consistent cut performance
- Three cut modes, all controlled by Instant Response™ technology, offer surgeons a variety of choices:
- Low Cut for delicate tissue or laparoscopic cases
- Pure Cut for a clean, precise cut
- Blend for cutting with hemostasis
- Four coag modes:
- Desiccate for low voltage contact coagulation suitable in laparoscopic and delicate tissue work
- Fulgurate (high crest factor) for efficient noncontact coagulation in most applications
- Fulgurate (low crest factor) for lower voltage coagulation requirements
- Spray for coagulating large tissue areas with superficial depth of necrosis
- Three bipolar modes:
- Precise, Standard, and Macrobipolar are controlled by the Instant Response™ system
- Precise and Standard setting utilize low voltage to prevent sparking
- Versatile system that is uniquely compatible with other devices, including:
- Force Argon™ II and Force GSU™ argon coagulation system
- CUSA EXcel™ and CUSA™ 200 ultrasonic surgical aspirators
- OptiMumm™ smoke evacuator, through a direct cable link
- Valleylab VLCM bipolar current monitor
- Compatible with and used as the electrosurgical energy source for:
- Dyonics* Control RF arthroscopic ablation system
- Dyonics* Electroblade™ rotary resection system
- Cook Vascular Perfecta™ EDS pacemaker lead extraction system
- Compatible with, and the exclusive electrosurgical generator for, the Computer Motion Hermes™ Voice Command system
Valleylab Force FX-C Specifications
Dimensions
- Height: 4.375 in ( 11.1 cm)
- Width: 14 in (35.6 cm)
- Depth: 18 in (45.7 cm)
- Weight: 18 lbs (<8.2 kg)
Operating Parameters
- Ambient Temperature Range: 10° to 40″ c (50° to 104° F)
- Relative Humidity: 30% to 75%, noncondensing
- Atmospheric Pressure: 700 to 1060 millibars
- Warm-up time: If transported or stored at temperatures outside the operating temperature range, allow one hour for the generator to reach room temperature before use.
Transport and Storage
- Atmospheric Temperature range: -40° to 70° C f40°:10 158° F)
- Relative humidity: 10% to 100%, condensing
- Atmospheric Pressure: 500 to 1060 millibars
- Duration of Storage: If stored longer than one year, the battery must be replaced (Refer to Battery Replacement in Section 7) and a full checkout, including calibration, must be completed before use.
Output Waveforms
- Bipolar
- Precise: 470 kHz sinusoid
- Standard: 470 kHz sinusoid
- Macro: 470 kHz sinusoid
- Monopolar Cut
- Low: 390 kHz sinusoid. Similar to the Pure Cut mode except the maximum voltage is limited to a lower value.
- Pure: 390 kHz sinusoid
- Blend: 390 kHz bursts of sinusoid, recurring at 27 kHz intervals. 50% duty cycle envelope.
- Monopolar Coag
- Desiccate: 240 kHz sinusoid repeated at 39 kHz. 8% duty cycle.
- Fulgurate: 390 kHz damped sinusoid bursts with a repetition frequency of 30 or 57 kHz into 500 ohms.
- Spray: 390 kHz damped sinusoidal bursts with a randomized repetition centered at 28 kHz. Frequencies include 21 kHzz < f < 35 kHz. Output is further modulated by a random 250 Hz envelope with a variable duty cycle.
- Output power changes by less than 15% or 5 watts, whichever is greater, as the line voltage varies from 104-132 volts and 208-264 volts (at rated load).
Low Frequency Leakage
- Source current, patient leads, all outputs tied together.
- Normal polarity, intact chassis ground: < 10 µA
- Normal polarity, ground open: < 50 µA
- Reverse polarity, ground open: < 50 µA
- Sink current, 140V applied, all inputs: < 50 µA
Input Power Requirements
- Operating range is 85 to 132 AC volts. Maximum current is 7 amperes in Cut and 4 amperes in Coag.
Rem Contact Quality Monitoring System
- Measurement Frequency: 80 kHz ± 10 kHz
- Measurement Current: Less than 10 µA
- Acceptable Resistance Ranges:
- REM™ pad – 5-135 ohms
- Non-REM™ pad – less than 20 ohms
- Acceptance range is 5-135 ohms after REM PolyHesive II return electrode is applied.
- Adaptive REM: REM trip is baseline impedance plus 40%. For example, if the baseline impedance is 30 ohms, the upper-level trip approximately 42 ohms. If the pad-patient impedance falls below the baseline impedance, a new baseline is established.