Description
The GE Aespire S5 Anesthesia Machine with 7900 SmartVent, available from PSM Medical Technologies, is a high-performance anesthesia solution designed to deliver precision, safety, and efficiency in the operating room. Engineered by GE Healthcare, this system combines the trusted reliability of the Aespire S5 platform with the advanced capabilities of the 7900 SmartVent, offering sophisticated ventilation modes for a wide range of patient conditions.
With features like electronic gas mixing, integrated patient monitoring, and a user-friendly interface, the Aespire S5 enhances workflow and patient care during anesthesia delivery. Whether for general surgery or critical care environments, PSM Medical Technologies supports this advanced anesthesia workstation with expert setup, training, and ongoing service.
SPECIFICATIONS
Physical Weight & Dimensions
- Height: 134.5 cm/52.9 in
- Width: 72 cm/28.3 in
- Depth: 73 cm/28.7 in
- Weight: Approximately 108 kg/238 lb
Top Shelf
- Weight limit: 34 kg/75 lb
- Width: 66 cm/26 in
- Depth: 40 cm/15.75 in
Work Surface
- Height: 81.7 cm/32.2 in
- Size: 2160 cm-2/334 in-2
DIN Rail
Side of machine: 34.5 cm/13.6 in
Drawers (internal dimensions)
- Height: 17.5 cm/6.9 in
- Width: 33 cm/13 in
- Depth: 26.5 cm/10.4 in
Absorber Bag Arm (optional)
- Arm length: 30.5 cm/12 in
- Bag arm height (adjustable): 87 cm (34.3 i) to 104 cm (40.9 in)
Casters
- Diameter: 12.5 cm/5 in
- Brakes: Individual locking front casters
Electrical Specifications
Current Leakage
- 100/120 V: < 300μA
- Power
- Power input: 100-120 Vac, 50/60 Hz
- Power cord: Length: 5 m/16.4 ft
- Rating: 15A @ 120 Vac
Inlet/outlet modules (120 V)
- System circuit breakers: 15A
- Outlets (optional): 4 outlets on back, 3-2A, 1-3A individual breakers, optional isolation transformer
Pneumatic Specifications
Auxiliary Common Gas Outlet
Connector: ISO 22 mm OD and 15 mm ID
Gas Supply
- Pipeline input range: 240 kPa to 600 kPa/35 psig to 88 psig
- Pipeline connections: DISS-male, DISS-female, DIN 13252, AS4059, S90-116 or NIST (ISO 5359). All fittings available for O2, N2O, and Air, and contain pipeline filter and check valve.
- Cylinder input: Pin indexed in accordance with CGA-V-1 or DIN (nut and gland); contains input filter and check valve
Note: Maximum 3 cylinders; two inboard mounted, one outboard mounted. - Primary regulator diaphragm minimum burst pressure: 2758 kPa/400 psig
- Primary regulator nominal output: < 338 kPa/49 psig Pin indexed cylinder connections
- < 407 kPa/59 psig DIN cylinder connections
Environmental Specifications
System Operation
- Temperature: 10° to 40°C/50° to 104°F
- Humidity: 15 to 95% relative humidity (non-condensing) per IEC 68-2-3
- Altitude: -440 to 3565 m/500 to 800 mmHg
System Storage
- Temperature: -15° to 50°C/-5° to 122°F
- Humidity: 10 to 95% relative humidity (including condensing) per IEC 68-2-3
- Altitude: -440 to 3565 m/500 to 800 mmHg
- Oxygen cell storage: -15° to 50°C/5° to 122°F; 10 to 95% relative humidity; 500 to 800 mmHg
Electromagnetic Compatibility
- Immunity: Complies with all requirements of EN 60601-1-2
- Emissions: CISPR 11 group 1 class B
- Approvals: UL 2601-1; CSA C22.2 #601.1; EN/IEC 60601-1; CE 0197