Maquet Datascope CS100
Maquet Datascope CS100
Maquet CS100 IABP Features
The Datascope SC 100 is an intra- aortic balloon pump that features Intellasync software. The Maquet Datascope CS100 and the Datascope Cs 100 are the same base system the only difference between them is the software on each unit. The CS100 has the more recently updated software. The Datascope IAPB can be used in an operating room, catheter lab, or critical care setting. The system offers three operation modes; automatic, semi-automatic, and manual, making it easy to switch between modes when needed. The Datascope CS100 offers a foldable and detachable laptop-style display. The color display screen makes it easier for caregivers to adjust the pump and monitor the patient remotely. Another version of the Maquet Datascope Intra-Aortic Balloon Pump is the Maquet Datascope CS 300.
- Maquet CS100 IABP comes with Intellisync and offers a one-button start-up, automatically adapting to changing conditions, making it the right choice for any patient condition.
- Automatically evaluates and selects optimal trigger source and lead
- Sets inflation at the dicrotic notch and deflates IAB at the start of systole
- Responds to changes in signal quality by selecting a new trigger source Automated trigger source and timing management to provide continuous, consistent support for your patient.
- Fast, one-button start-up automatically conducts autofill and initiates 1:1 assist.
- Full color, high contrast display tilts for easy viewing.
- Display and keypad in laptop-style configuration, allowing caregivers to fold and detach the unit for remote operation. Easy-to-use keypad with an intuitive user interface.
- Customizable display panel allows operators to adjust sweep speed, brightness, balloon waveform, and inflation markers.
- Switch easily between three operation modes: Automatic, Semi-Automatic, and Manual.
- Automatically evaluates and selects optimal trigger source and lead.
- System detects changes and easily adapts to new patient rhythms.
- Fully annotated printouts with user-configurable waveforms and strip lengths.
- Locking casters and durable design keep the CS100 secure in a variety of settings.
Maquet CS100 IABP Specifications:
Dimensions
- Height: 30 in (76.2 cm)
- Width: 22.5 in (57.2 cm)
- Depth: 13.2 in (33.5 cm)
- On Cart
- Height: 42.2 in (114.8 cm)
- Width: 17.6 in (44.7 cm)
- Depth: 23.4 in ( 59.4 cm)
- Off Cart
- Height: 28.2 in (71.6 cm)
- Width: 11.3 in (28.7 cm)
- Depth: 21.5 in ( 54.6 cm)
- Monitor weight: 9.5 lb (4.3 kg)
- Console weight: 133.2 lb ( 60.4 kg)
Color Display
- Width:8.3 in (21cm)
- Height: 6.2 in (15.8 cm)
- viewing angle:
- Up: 45°
- Down 55°
- Right 70°
- Left 70°
- Rotates 330°
- Tilts180°
- Detachable Laptop-like closure for storage and protection
Power
- Mains Voltage: 100-120 VAC ± 10% or 220-240 VAC ± 10%
- Mains Frequency: 50/60 Hz ± 3 Hz
- Internal Battery: 24 VDC (nominal), 17.2 Amp-hour, 3 hrs.(approx. @ 90 bpm)
- Battery Type: Maintenance-free; Sealed lead-acid
General information
- ECG Gain (default): 1 V output per 1 mV input ±5% (waveform automatically scaled to occupy ECG display window)
- Gain (variable): 0.15 to 3.0 cm/mV ±20% (autoscaling disabled)
- Frequency Response: 0.5-12 Hz (display); 0.5-135 Hz (Output to Monitor)
- Defibrillator Protection: Discharge levels 360 J (trace returns to the screen in 5 sec. max)
Operating Conditions
- Operating Temperature: 10°C - 40°C
- Operating Humidity: 5 - 95% (R.H.) non-condensing
- Operating Altitude: 0 - 12,000 feet (3,657 m); automatic altitude correction for IAB pressure
Operating Modes
- ECG Trigger: Threshold dynamically adjusted by a system for high sensitivity and selectivity of the R-wave detection; Minimum = 120μV ± 20μV at normal gain; 40μV at max. gain
- Pressure Trigger: Default trigger threshold is automatically adjusted to 38% of the difference between peak systolic and end-diastolic (avg. over multiple cycles); In variable mode: User-adjustable between 7 and 30 mmHg ± 3 mmHg
- Pacer A Trigger: R-wave detection (as above) except pacer blanking is extended to 100 ms
- Pacer V/A-V Trigger: V Pacer: fixed at a rate up to 185 bpm (no demand pacing) A-V Pacer: fixed at a rate up to 125 bpm (no demand pacing) with A-V intervals between 80-224 ms
- Internal Trigger: Variable mode: 40-120 bpm; Normal mode: 80 ± 1 bpm
- Tall T-Wave Rejection: Rejects all T-Waves where the Q-T interval is <300 ms and the amplitude is <70% of QRS input amplitude
- Pacer Rejection: Rejects all pulses of amplitude ± 2.0 mV to ± 700 mV and durations between 0.1 ms to 2.0 ms with 1) No tail; 2) 100 ms time constant tail < 1 mV; 3) 25 ms time constant tail < 1 mV; 4) 4 ms time constant tail < 2 mV